Cat. No. | Model | Spectral Bandwidth | Cuvette |
---|---|---|---|
S23-3269-030 | OPTIZEN NanoQ | 1.0 nm (FWHM at Hg 253.7 nm) | X |
S23-3269-035 | OPTIZEN NanoQ Plus | O |
Cat. No. | S23-3269-030 | S23-3269-035 | |
---|---|---|---|
Model | OPTIZEN NanoQ | OPTIZEN NanoQ Plus | |
Photometrics System | Microvolume Spectrophotometer | ||
Light Source | Xenon flash lamp | ||
Spectral Bandwidth | 1.0 nm (FWHM at Hg 253.7 nm) | ||
Wavelength Range | 190 to 850 nm | ||
Wavelength Display(setting) | 1 nm | ||
Wavelength Accuracy | ¡¾ 1 nm | ||
Life Time | Up to 10 years | ||
Detector | CMOS linear image sensor (2048 pixels) | ||
Microvolume | Minimum Sample Volume | 1 ¥ìL | |
Photometric Range | 0.02 - 330 A (10 mm equivalent) | ||
Detection Limit | 2 ng/¥ìL (dsDNA) 0.06 mg/mL (BSA) 0.003 mg/mL (IgG) |
||
Maximum Concentration | 16,500 ng/¥ìL (dsDNA), 400 mg/mL (BSA) | ||
Cuvette | Photometric Range | - | 0 - 2 A |
Detection Limit | - | 0.2 ng/¥ìL (dsDNA) 0.006 mg/mL (BSA) 0.0003 mg/mL (IgG) |
|
Center Height (Z-height) | - | 15 mm | |
Heating (Optional) | - | 37 ¡ÆC | |
Absorbance Precision | 0.002 A (0.5 mm path) or 1% | ||
Absorbance Accuracy | 3% (at 0.97A at 302 nm) | ||
Measurement Time | < 8 seconds | ||
Display | 7 inch color LCD with touch screen | ||
Operating System(OS) | Android¢â | ||
Software Compatibility | Windows 7 and 10 | ||
Weight(kg) | 3 kg | ||
Footprint(W£øD) | 216 x 290 mm |